FDAApril 22, 2020device

ARTIS Pheno - Interventional Fluoroscopic X-Ray System Model: 10849000 - Product Usage: is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

If the C-arm leaves its intended travel path due to a fault within the drivetrain, movements may be impacted or impossible and the area of interest cannot be reached. In the event the C-arms movements are not possible, the system can only be returned to normal operation with the support of a field service engineer

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical