FDAJanuary 7, 2026device

See full list in RES, exceeds character limit 1) GYN LAPAROSCOPY CDS, Model Number: CDS981504G; 2) BAP NORTH ROBOTIC SURGERY CDS, Model Number: CDS982407P; 3) ROBOTIC GYN CDS-1, Model Number: CDS982893I; 4) GYN LAPAROSCOPY CDS, Model Number: CDS983300J; 5) NHP ROBOTIC GYNE CDS, Model Number: CDS9...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

What to do

FDA enforcement status: Ongoing

Brands named

medline industries lpmedlinemedline industries

UPCs

101953271579684019532715796910195327110116401953271101171088994220805040889942208051101934894250484019348942504910193489872521401934898725221019532707232240195327072323

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.