FDAFebruary 15, 2012device

Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A Discrepancy was found on Hitachi AST (HA7 61) package insert. Limitations section indicates linearity is 500 U/L and the Performance section linearity is listed as 0-600 U/L, which is not consistent with the 510k submission package insert.

What to do

FDA enforcement status: Terminated

Brands named

medtest

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Pointe Scientific Hitachi AST Reagent R1 : PE plastic bottle w/PP plastic cap. R2: PE plastic bottle w /PP plastic cap. Product Usage: Two Part Chemistry Reagent For the quantitative determination of Aspartate Aminotransferase (ASl) in human serum on Hitachi analyzers. — Recall Details · AllClear