FDAApril 6, 2018device

Non-Locking Screw, D 2.7MM X 8MM, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size (Part Numbers): a) D 2.7MM X 8MM (002-D-27008); b) D 2.7MM X 10MM (002-D-27010); c) D 2.7MM X 12MM (002-D-27012); d) D 2.7MM X 14MM (002-D-27014); e) D 2.7MM X 16MM (002-D-27016); f) D 2.7MM X 18MM (002-D-270...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Firm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.

What to do

FDA enforcement status: Terminated

Brands named

orthofix

UPCs

80560996457821805609964579918056099645805180560996458121805609964582918056099645836180560996458431805609964585018056099645867180560996458741805609964588118056099645898