FDAMay 26, 2016device

Medtronic, MyCareLink Patient Monitor, Model 24950, Rx Only. The MyCareLink Monitor, Model 24950, is a remote monitoring system that interrogates implanted devices and transmits the data to the Care Link Network for viewing by the physician.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Recently, a new software version was automatically sent to a subset of Model 24950 MyCareLink Monitors. After release, Medtronic identified an issue with the software that prevents implanted device data from being available to clinicians on the CareLink Network. While the transmission appears successful to the patient, the transmitted data, including CareAlerts, are not visible to the clinic.

What to do

FDA enforcement status: Terminated