FDAApril 17, 2020device

Cios Spin Image-Intensified Fluoroscopic X-Ray System, Model Numbers VA30-10308194

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The unplugging of the main cable on the monitor trolley from the X10 connector can cause an unsafe electrical voltage (up to 67,6 V DC) at the X10 connector. This can result in an electrical shock if the user or other persons touch the contact pins of the X10 connector.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical