FDAMay 8, 2023device

VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of transferrin, C3, C4, IgA, IgG and Ig...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Potential to cause biased results in the upper end of the reportable range for the quantitative measurement of Transferrin, C3, C4, IgA, IgG and IgM. Within the reference interval, the highest bias observed in selected sample testing investigation was +35.3% for IgM; -4.3% for IgG; +12.3% for IgA; -6.9% for C3; -6.6% for C4; and -17.8% for transferrin. Falsely elevated or reduced results could be reported depending on the analyte and/or sample concentration.

What to do

FDA enforcement status: Ongoing

Brands named

ortho clinical diagnosticsorthoortho clinical

UPCs

10758750006687

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
VITROS Chemistry Products Calibrator Kit 20-in vitro diagnostic use only. VITROS Chemistry Products Calibrator Kit 20- Used to calibrate the VITROS 5,1 FS/4600 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems for the quantitative measurement of transferrin, C3, C4, IgA, IgG and Ig... — Recall Details · AllClear