FDAMay 17, 2024device

Stimuplex A, 30 DEG, 24GX1", 0.55x25mm, Catalogue Number: 4894251.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The devices are labeled "not made with DEHP" or the DEHP free symbol; however, the glue used contains traces of DEHP.

What to do

FDA enforcement status: Ongoing

Brands named

b braun medical