FDAJune 12, 2015device
ATTUNE FB PS ARTICULATION SURFACE SIZES 3-8 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a c...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.
What to do
FDA enforcement status: Terminated
Brands named
depuy orthopaedicsdepuy
UPCs
106032951324311060329513244810603295132455106032951324621060329513247910603295132486
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 7 CEM. Part Number: 150450107.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 5 CEM. Part Number: 150450205.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 3 CEM. Part Number: 150450203.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 4 CEM. Part Number: 150450204.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 5 CEM. Part Number: 150450105.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 4 CEM. Part Number: 150450104.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 7 CEM. Part Number: 150450207.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 8 CEM. Part Number: 150450208.2026-04-03