FDAOctober 16, 2018device

MED EL Cochlear Implant System, PULSARci100 Standard - Product Usage: The PULSARci100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sen...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.

What to do

FDA enforcement status: Terminated

Brands named

med el elektromedizinische gereatemedmed el

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.