FDAMarch 8, 2013device

LifeCare PCA infusion system with Hospira MedNet software allows clinicians to administer, or patients to self-administer analgesia, safely and effectively within clinician programmed limits and/or hospital-defined medication limits. The LifeCare PCA infusion system is used in a wide range of cli...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.

What to do

FDA enforcement status: Terminated

Brands named

hospira

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
LifeCare PCA infusion system with Hospira MedNet software allows clinicians to administer, or patients to self-administer analgesia, safely and effectively within clinician programmed limits and/or hospital-defined medication limits. The LifeCare PCA infusion system is used in a wide range of cli... — Recall Details · AllClear