FDAMay 16, 2024device

US AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.

What to do

FDA enforcement status: Ongoing

Brands named

fresenius medical carefreseniusfresenius medical

UPCs

04251285200357

Recall history