FDAJune 5, 2023device
Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers 0998-00-0800-75 0998-00-0800-83 0998-00-0800-85
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
IABP may lose the ability to charge batteries in one or both bay slots. Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC power. Low battery alarms may alert the User to the issue prior to interruption of therapy.
What to do
FDA enforcement status: Ongoing
Brands named
datascope
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABattery Charging Station; Model: 0998-00-0802;2026-02-06
- FDACS100 IABP. Software Version CS100 IABP Q.01.2026-01-23
- FDACS300 IABP. Software Version CS300 IABP C.01.2026-01-23
- FDACardiosave Rescue. Intra-Aortic Balloon Pump system.2025-11-04
- FDACardiosave Hybrid. Intra-Aortic Balloon Pump system.2025-11-04
- FDACardiosave Hybrid2025-10-23
- FDACardiosave Rescue2025-10-23
- FDAMEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-012024-09-17