FDAMay 10, 2024device

Xstar Safety Slit Knife, 2.75 mm 45¿ Bevel Up, Model Number 378227, a component of CustomEye Kits, Numbers 584562, 5800109

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The wrong configuration of the blade was in the package. The bevel was on the opposite side, instead of being in front (bevel up) it was in the back (bevel down).

What to do

FDA enforcement status: Ongoing

Brands named

beaver visitecbeaver

UPCs

00886158001652