FDAApril 12, 2017device
SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device which support the heart's left ventricle by increasing coronary perfusion and reducing the work of the left ventricle.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.
What to do
FDA enforcement status: Terminated
Brands named
datascope
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDABattery Charging Station; Model: 0998-00-0802;2026-02-06
- FDACS100 IABP. Software Version CS100 IABP Q.01.2026-01-23
- FDACS300 IABP. Software Version CS300 IABP C.01.2026-01-23
- FDACardiosave Rescue. Intra-Aortic Balloon Pump system.2025-11-04
- FDACardiosave Hybrid. Intra-Aortic Balloon Pump system.2025-11-04
- FDACardiosave Hybrid2025-10-23
- FDACardiosave Rescue2025-10-23
- FDAMEGA 7.5Fr. 40cc Intra-Aortic Balloon with Accessories, Model Number 0684-00- 0295-012024-09-17