SHBG IMMULITE 2000/IMMULITE 2000 XPi For the in vitro diagnostic use with the IMMULITE 2000 Systems Analyzers-For the quantitative measurement of SHBG in serum, as aid in the differential diagnosis of hirsutism.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Certain lots of Sex Hormone Binding Globulin (SHBG) do not meet the thirty (30) day open vial adjustor stability claim when reconstituted and stored at 2-8¿C, as published in the Instructions for Use (IFU). Adjustors stored at 2-8C after 14 days may show a bias in patient samples and quality control falling outside of published ranges. Lookback not recommended.

What to do

FDA enforcement status: Terminated