FDAApril 5, 2024device

CARESCAPE B850 Model Numbers: a) MBC303, Model Numbers: 1) 2078633-001, 2) 2078633-010, 3) 2078633-012, 4) 2078633-017, 5) 2078633-028, 6) 2095802-001-54069330, 7) 2095802-001-554922, 8) 2095802-001-554923, 9) 2095802-001-554924, 10) 2095802-001-554925, 11) 2095802-001-554926, 12) 2095802-001-554...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.

What to do

FDA enforcement status: Ongoing

Brands named

ge healthcare finland oy

UPCs

00840682136723