FDAJuly 14, 2025device

Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The

What to do

FDA enforcement status: Ongoing

Brands named

baxter healthcarebaxter

UPCs

05413765851797

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.