FDAJune 30, 2021device

Kaguya Automated Peritoneal Dialysis System

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cyclers Extra Last Drain Mode. In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.

What to do

FDA enforcement status: Terminated

Brands named

baxter healthcarebaxter