FDAJune 6, 2016device
Target Detachable Coils: TARGET HELICAL NANO 2MM X 6CM MODEL Number:M0035452060 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
What to do
FDA enforcement status: Terminated
Brands named
stryker neurovascularstryker
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDANeptune Smart Docking Station (120V), Catalog Number 0770-100-000, part of the Neptune Waste Management System2026-05-06
- FDADaig Livewire Steerable, Product Number 401582; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic XT Steerable, Product Number 201104; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic Deca Steerable, Product Number 201101; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABARD Dynamic Tip Steerable, Product Number 6DYNTP001;2026-04-10
- FDABARD Dynamic Tip Steerable, Product Number 200344; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDABW Webster Duo-Decapolar, Product Number D728260RT; REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
- FDADaig Livewire Steerable, Product Number 401603, REPROCESSED ELECTROPHYSIOLOGY CATHETER2026-04-10
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →