Dimension Gentamicin Flex reagent cartridge, DM GENT, gentamicin enzyme immunoassay, Lot numbers affected are EB8115, BB8159, and FB8275.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The affected lots may exhibit inaccuracy for Quality Control and patient samples at the low end of the Analytical Measurement Range. The bias is due to the calibration curve not showing separation between the Level 1 and Level 2 calibrators.

What to do

FDA enforcement status: Terminated