FDAJune 3, 2021device

HeartWare HVAD Pump Implant Kit, REF 1125

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart.

What to do

FDA enforcement status: Ongoing

Brands named

heartware

UPCs

008887070002220088870700764100888707005685