FDAJune 16, 2016device

Medtronic, Temporary Pacing Lead System, Model 6416, Sterile EO. Product Catalog Numbers: Five (5) Pack Single Pack 6416-100 6416-1005 6416-140 6416-1405 6416-200 6416-2005 For temporary intracardiac pacing and EGM recording. The system is disposable, for temporary single patient use, with a cont...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

This Pacing Lead System does not comply with Section 8.5.2.3 of IEC 60601-1, and corresponding provisions of FDA 21 CFR 898, which relate to design standards to prevent connecting a patients lead to a possible hazardous voltage.

What to do

FDA enforcement status: Terminated

Brands named

medtronic inc cardiac rhythm and heart failuremedtronicmedtronic inc