Swan-Ganz Catheter, Models: 131F7/131F7P, 131F7J, 141F7, 151F7, 834F75, 096F6/096F6P, C144F7, 782F75M, 132F5, C146F7, 831F75/831F75P, TS105F5, 774F75, 777F8, AIQSGF8;

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Catheters may, due to a manufacturing process and material changes, leak/break from the (blue) proximal injectate lumen hub, which may lead to infection, medication loss, and/or blood loss.

What to do

FDA enforcement status: Ongoing

Brands named

becton dickinson andbectonbecton dickinson

UPCs

006901030004120069010300040500690103000504006901030005660069010315175600690103000245006901032186020069010300513400690103197303006901030004670069010315313200690103218282