FDAJuly 19, 2019device

NobelActive Internal NP 3.5x15mm Endosseus dental implant, REF: 34128, Rx Only, Sterile R, CE, UDI: (01)07332747010619

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.

What to do

FDA enforcement status: Terminated

Brands named

nobel biocarenobel

UPCs

07332747010619