FDAJune 23, 2023device

MEGA 8Fr. IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number MEGA 8Fr. 50cc IAB WITH STATLOCK D684-00-0296-01 MEGA 8Fr. 50cc IAB WITH STATLOCK & APA D684-00-0296-02 MEGA 8Fr. 50cc IAB WITH ACCESSORIES (W/O STA...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

What to do

FDA enforcement status: Ongoing

Brands named

datascope

UPCs

1060756710800110607567108018106075671094111060756711326510607567113272106075671096191060756710962610607567107301106075671097561060756710976310607567107493