FDAJune 23, 2023device

LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Description Part Number LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES D684-00-0479-01 LINEAR 7.5Fr. 34cc IAB WITH ACCESSORIES (APA) D684-00-0479-02 LINEAR 7.5Fr. 34cc IAB WITH...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

What to do

FDA enforcement status: Ongoing

Brands named

datascope

UPCs

106057671065571060576710654010605767109329106057671093361060576710969510605767109701106057671131801060576711314210605767109503106057671095101060756710656410607567106571