FDAJanuary 12, 2021device

Grasper/Retriever, Alligator Jaw, Rat Tooth 2.4 mm x 8 mm x 230 cm Catalog Number: 7226 Intended use: single use accessory devices used through the working channel of a GI endoscope

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Mislabeled: Devices labeled as catalog #7208 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm) may actually be catalog #7226 (Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 8 mm x 230 cm) and Devices labeled as catalog #7226, (Grasper/Retriever, Alligator Jaw, Rat Tooth, 2.4 mm x 8 mm x 230 cm) may actually be catalog #7208, Grasper/Retriever, Alligator Jaw, 2.4 mm x 5 mm x 230 cm).

What to do

FDA enforcement status: Completed

Brands named

shent