FDAJune 5, 2023device

Cios Select VA 21 with FD- A Mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. Material #11515088

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Incorrect glue was applied on the plexiglass sheet on the single tank cover assembly could potentially fall off and the ingress of liquid may occur, the risk of electric shock exists (e.g., ingress of a large amount of liquid following the plexiglass dropping off, at the same time the user or operator happens to touch the liquid on the cover which is connected to the internal live parts).

What to do

FDA enforcement status: Ongoing

Brands named

siemens medical solutionssiemenssiemens medical

UPCs

04056869148243

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.