Computed Tomography X-ray systems Brilliance iCT, Philips Healthcare System. Intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict

What to do

FDA enforcement status: Terminated

Brands named

philips medical systems clevelandphilipsphilips medical