FDAJune 22, 2016device

Olympus Scientific Solutions Americas Corporation (OSSA) Model: Analytical X-Ray System X -Ray Fluorescence. Intended to emit ionizing radiation Delta with models: DC-2000, DC-4000, DC-6000, DI-2000, DP-2000, DP-4000, DP-4050, DP- 6000, DPO-4000, DPO-6000, DS-2000, DS-4000, DS-6000, DS-6500, De...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.

What to do

FDA enforcement status: Ongoing

Brands named

olympus scientific solutions americasolympusolympus scientific

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.