FDAMay 24, 2023device
NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems (NeuMoDx System(s) Ref: 100100
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Top label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results
What to do
FDA enforcement status: Ongoing
Brands named
neumodx molecularneumodx
UPCs
10814278020274
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAAlinity m Resp-4-Plex Amplification (AMP) Kit (CE), List Number 09N79-090, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the qualitative detection and differentiation of RNA from influenza A virus (flu A), influenza B virus (flu B), Respiratory...2025-04-28
- FDAAlinity m Resp-4-Plex AMP Kit, List Number 09N79-096, a multiplex real-time reverse transcription (RT) polymerase chain reaction (PCR) test intended for the simultaneous qualitative detection and differentiation of RNA from SARS-CoV-2, influenza A virus (flu A), influenza B virus (flu B), and/or ...2025-04-28
- FDAThe Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).2025-03-20
- FDAAbbott Alinity m HPV AMP Kit, used with the Alinity m System, product codes: a) REF 09N15-095; b) REF 09N15-090; c) REF 09N15-0912024-09-12
- FDAAbbott Alinity m STI AMP Kit, used with the Alinity m System, product codes: a) REF 09N17-095; b) REF 09N17-090; c) REF 09N17-091;2024-09-12
- FDANxTAG Respiratory Pathogen Panel +SARS-CoV-2, REF: I056C04682024-09-05
- FDANxTAG Respiratory Pathogen Panel, REF: I051C04472024-09-05
- FDAAlinity m System, Part Number: 08N53-0022024-03-20
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.