ProxiDiagnost N90, UDI 706100

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The thermo switches in the three-phase transformer, which normally aid in powering down the System Power Distribution Unit, may be incorrectly installed and not working. If the transformer overheats due to a first failure and the thermo switch is not activated, the device may begin to generate smoke.

What to do

FDA enforcement status: Terminated

Brands named

philips north americaphilipsphilips north