FDAApril 25, 2023device
Incisive CT Plus-Computed Tomography X-Ray System Model: 728149
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Metal mounting box on the rotating scanner on rotor (heat exchanger box) located within the Incisive CT system may become expelled and make contact with other components and cause injury to user or if detached a loud noise will be emitted and the system will shut down
What to do
FDA enforcement status: Ongoing
Brands named
philips north americaphilipsphilips north
UPCs
00884838103474
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- CPSCGreenworks Tools Recalls 24V and 48V Kobalt Yard Power Tools with USB-C Batteries Due to Risk of Serious Injury from Fire Hazard2026-07-09
- FDAPhilips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.2026-06-05
- FDAPhilips Avalon Fetal Monitor, FM 30 Part numberM2703A2026-05-26
- FDAPhilips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A2026-05-26
- FDAPhilips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # ...2026-05-18
- FDAPhilips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Elev...2026-05-18