FDAApril 8, 2020device

therascreen EGFR RGQ PCR Kit (24), Reference Number REF 870121

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a risk for a false mutation positive result in rare cases resulting from a fluorescence artefact. In these rare cases, the controls pass correctly, while a fluorescence artefact in the mutation assay causes an incorrect valid mutation positive result.

What to do

FDA enforcement status: Terminated

Brands named

qiagen sciencesqiagen

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.