FDAMay 10, 2017device

Vitality Spinal Fixation System, Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs Model Number(s): 07.02066.001, 07.02053.001, 730M0017

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Zimmer Biomet is conducting a medical device recall for Vitality T27 Final Drivers and Vitality Torque Limiting Handle  90 in-lbs because when applying torque to tighten the closure top the driver has the potential to create a torsional spring which can cause damage and the potential for the values of torque to go out of specification.

What to do

FDA enforcement status: Terminated

Brands named

zimmer biomet spinezimmerzimmer biomet