FDAJuly 2, 2019device

QIAsymphony SP SOW 5.0.3, software used with the QIAsymphony SP Instrument (The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids-IVD) Model Number: 9001297 Product Usage: The QIAsymphony SP instrument is designed to perform automated purification of nucleic...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

QIAsymphony software version 5.0.3 - Software issue incorreclty assigns the 2D bar code eluate ID to the sample within the results file. If 2D bar code eluate IDs are not checked against the sample ID prior to downstream application the incorrect tube may be selected and it could have the potential to lead to delayed results or incorrect patient reporting

What to do

FDA enforcement status: Ongoing

Brands named

qiagen sciencesqiagen

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
QIAsymphony SP SOW 5.0.3, software used with the QIAsymphony SP Instrument (The QIAsymphony SP instrument is designed to perform automated purification of nucleic acids-IVD) Model Number: 9001297 Product Usage: The QIAsymphony SP instrument is designed to perform automated purification of nucleic... — Recall Details · AllClear