FDAMay 27, 2021device
Cytocell LPH019-A E2A (TCF3) Breakapart Probe- In Vitro Diagnostic Analyte specific reagents (ASR's)
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
May show unexpected locus specific signals in addition to those at 19p13.3. Users may observe faint additional locus specific signals at 9q34
What to do
FDA enforcement status: Terminated
Brands named
cytocell
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDARET Proximal Probe. Used in Fluorescence in situ hybridization (FISH), a laboratory technique for detecting and locating a specific DNA sequence on a chromosome.2023-05-24
- FDARET Distal Probe. Used in Fluorescence in situ hybridization (FISH), a laboratory technique for detecting and locating a specific DNA sequence on a chromosome.2023-05-24
- FDACytoCell BCL11B Proximal in FITC Spectrum-LDT for investigation T-cell related malignancy. This device is an analyte specific reagent probe covering the 14q32.2 locus proximal to the BCL11B gene on chromosome 14. Ref: MPD397812022-10-27
- FDACytoCell BCL11B Distal in Texas Red Spectrum- Analyte Specific Reagent Ref: MPD397822022-10-27
- FDACytocell 8 Square Template Slides (Glass)- microscope slides Reference: PCN0082022-02-14
- FDATCL1 Breakapart Probe, Fluorescence in situ hybridisation (FISH) is a technique allows the visualisation of DNA sequences upon chromosome Catalog Number: LPH046-A2022-02-04
- FDADel(5q) Deletion FISH Probe Kit: In-Vitro Diagnostic to detect deletions in chromosome location from patients with AML or MDS Catalog Number: USA-LPH0242022-01-18
- FDACytoCell DiGeorge/VCFS TUPLE1 Region and 22q13.3 Region Probe, Model Number LPU004-A. analyte specific reagent for in vitro diagnostics2021-05-17
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.