FDAJune 3, 2025device

Hamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.

What to do

FDA enforcement status: Ongoing

Brands named

hamilton medical aghamiltonhamilton medical

UPCs

07630002808590

Recall history