FDAApril 16, 2018device

Merlin PCS Programmer Software: 3330 The Merlin PCS programmer (model 3650) is a portable, external device used to support the implanted device, and for this recall the software that resides on the programmer (software model 3330) is the method of delivery of the new firmware.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.

What to do

FDA enforcement status: Terminated