VITROS Chemistry Products Ca Slides; Catalog Number: 145 0261;
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Generations 67 and above of VITROS Ca Slides are experiencing an increased occurrence of Condition Code TH4-63J during calibration on VITROS XT 3400 and VITROS XT 7600 Systems, causing calibration failures that delay calcium test results and potentially impact patient management. The imaging reflectometer algorithm detects excessive "spikes" (imperfections) in slide images and generates the TH4-63J condition code when the threshold of 600 spikes is exceeded. Updated analysis revealed a 19-fold increase in failure rates from 0.0043% (2024 baseline) to 0.082% for Generation 67+ slides in 2025, with 100% of calibrator failures occurring at the lowest concentration level. This can result in delayed diagnosis and treatment of conditions such as hypercalcemic crisis, parathyroid disorders, and other calcium-related emergencies requiring urgent intervention.
What to do
FDA enforcement status: Ongoing
Brands named
UPCs
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDASiemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay2026-05-07
- FDAdirectCHECK ACT-LR whole blood quality control, level 2 Model/Catalog Number: DCGLR-22026-04-30
- FDABioFire Joint Infection (JI) Panel REF: RFIT-ASY-0138, 30 test Kit.2026-04-22
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;2026-04-02
- FDABiofire Spotfire Respiratory/Sore Throat Panel REF: 4234852026-04-01
- FDADimension Creatinine Flex reagent cartridge. Material Number: 10872079.2026-03-18
- FDALeadCare Ultra Blood Lead Test Kit; Catalog Number: 70-8098; The LeadCare Ultra kit is provided with the following materials: " Sensors (8 containers of 24 each; 192 tests) " Treatment Reagent Tubes (8 containers of 24 each; 192 tests) " Calibration Button (1 @) " Lead Control Level 1 (1 @ 2 mL) ...2026-03-13
- FDA(1)(2) LeadCare¿ II Blood Lead Test Kit, (3)LeadCare¿ II 5X Blood Lead Test Kit, (4)LeadCare¿ II 9X Blood Lead Test Kit ; Catalog Number(s): (1) 70-6762, (2) 70-6762RUO, (3) 70-8404 (Box containing 5 Test Kits), (4) 70-9000 (Box containing 9 test kits); The LeadCare II kit is provided with the fo...2026-03-13