FDAJuly 25, 2025device
Electrosurgical, Cutting and Coagulation and Accessories. Electrosurgical Generator ESG-410. Model Number: WA91307C. The ESG-410 Electrosurgical Generator is an electrosurgical generator which utilizes monopolar and bipolar high frequency current and supports ultrasonic instruments during open su...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Potential for the ESG-410 Electrosurgical Generator to display an error message: "E0662 System Error", followed by either automatic single reboots of the device, or continuous reboot 'loops'.
What to do
FDA enforcement status: Ongoing
Brands named
olympus
UPCs
04042761086349
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2113.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21122026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2118.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2110. May be included with the Olympus OER-ELITE.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-21112026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2117.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2115.2026-03-25
- FDAOER-ELITE Endoscope Reprocessor Connecting Tubes. Model/Catalog Number: MAJ-2116.2026-03-25
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →