VITEK¿ 2 Gram Positive Susceptibility card (AST-GP67), REF 22226, 20 cards per carton.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).

What to do

FDA enforcement status: Terminated

Brands named

biomerieux

UPCs

132016510310132017210317132017220317132017920324132018610331132019310307132019410308132021510329132021520329132022210306132022310307132024210326