FDAFebruary 2, 2015device

Invacare¿ Pronto¿ Air Personal Transporter (power wheelchair), Invacare Corporation, One Invacare Way, Elyria, Ohio, 44035-4190, (800) 333 -6900.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The batteries used in the Invacare¿ Pronto¿ Air Personal Transporter models power wheelchairs have the possibility to overheat during charging which could potentially result in leaking battery acid and/or swelling battery, loss of use, chemical burns and/or property damage.

What to do

FDA enforcement status: Terminated

Brands named

invacare