FDAJuly 11, 2016device

Alere INRatio/ INRatio2 PT/INR Test Strips Model Number: 99007EU, 99007G1, 99007G3, 99007G5, 99007G7, 99008EU, 99008G1, 99008G3, 99008G5, 99008G7 Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Alere San Diego is initiating a voluntary recall for the Alere INRatio¿/INRatio¿2 PT/INR Test Strips. This action is being initiated as a result of patients who had a therapeutic or near-therapeutic INR with the Alere INRatio¿/INRatio¿2 PT/INR System but a significantly higher INR (outside of therapeutic range) when performed by a central laboratory.

What to do

FDA enforcement status: Terminated

Brands named

alere san diegoalerealere san

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Alere INRatio/ INRatio2 PT/INR Test Strips Model Number: 99007EU, 99007G1, 99007G3, 99007G5, 99007G7, 99008EU, 99008G1, 99008G3, 99008G5, 99008G7 Product Usage: The Alere INRatio¿/INRatio¿2 system uses a modified version of the one-stage PT test. After a drop of blood is applied to the test strip... — Recall Details · AllClear