FDAMay 11, 2020device

VITROS Chemistry Product AMYL Slides Reagent, AMYLASE/5 PACK/90 SLDS, Product Code 8112724, UDI # 10758750004393 - Product Usage: For in vitro diagnostic use only to quantitatively measure amylase activity in serum, plasma, and urine using VITROS 250/350/5, 1 FS/4600/XT 3400 Chemistry Systems and...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Multiple coatings of VITROS Chemistry Products AMYL Slides are showing imprecision on control fluids when compared to the expected within lab SD as listed on the Performance Verifier (PV) Assay Sheets.

What to do

FDA enforcement status: Terminated

Brands named

ortho clinical diagnosticsorthoortho clinical

UPCs

600203565250600203565563600203565573600203565902600203566266600203566278600203566676600203567387600203567388600203567389600303570323600303570338