FDAJune 28, 2016device

Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Synthes 2.4mm Titanium Locking Screws Self-Tapping with Stardrive Recess 12mm, Plate, Fixation, Bone, Orthopedic Devices The 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm is a part of the followi...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

It was reported that the specified lots of the 2.4mm Titanium Locking Screw Self-Tapping with Stardrive Recess 12mm can break between the head and the shaft. This resulted from the neck feature being too close to the StarDrive feature leaving a thin wall between neck and drive.

What to do

FDA enforcement status: Terminated

Brands named

synthes usa productssynthessynthes usa