Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.