FDAJune 23, 2023device
ellacor System with Micro-Coring Technology-Indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. Model #: Gen 4
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring
What to do
FDA enforcement status: Ongoing
Brands named
cytrellis biosystemscytrellis
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAHistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.2025-04-02
- FDABrand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.012025-01-30
- FDALeica Biosystems Cryostat Model Number CM1950, Material Numbers: a) 14047742467, b) 1491950C1US, c) 1491950C2US, d) 1491950C3US; for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM1850, Material Numbers: a) 14047131148, b) 14901850001, c) 14901850001R0501, d) 1491850UVUL; for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM1900, Material Numbers: a) 14045227382, b) 14901900100, c) 14901900700, d) 1491900UVU2; for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples2024-09-04
- FDALeica Biosystems Cryostat Model Number CM3050 S, Material Numbers: a) 14047033518, b) 14903050S01, c) 14903050S02, d) 14903050S03; for freezing and sectioning tissue samples2024-09-04
Own something like this?