FDAAugust 10, 2021device
TRUMATCH CT PIN GUIDE FEM R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420581
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
What to do
FDA enforcement status: Terminated
Brands named
depuy orthopaedicsdepuy
UPCs
10603295430773
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 7 CEM. Part Number: 150450107.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 5 CEM. Part Number: 150450205.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 3 CEM. Part Number: 150450203.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 4 CEM. Part Number: 150450204.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 5 CEM. Part Number: 150450105.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL LT SZ 4 CEM. Part Number: 150450104.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 7 CEM. Part Number: 150450207.2026-04-03
- FDADepuy Synthes, ATTUNE REVISION HINGE FEMORAL RT SZ 8 CEM. Part Number: 150450208.2026-04-03